MODEL RANGE | |
---|---|
PLH120-P | 0 to 120V at 0 to 750mA |
PLH250-P | 0 to 250V at 0 to 375mA |
DIGITAL BUS INTERFACES | |
---|---|
Full remote control and read-back using RS-232, USB or LAN (LXI-C). All interfaces are at ground potential and opto-isolated from the outputs. Note: Remote/Local Sense is manually selectable only. | |
RS-232 | Standard 9-pin D connector. Baud rate 19,200 max |
USB | Standard USB hardware connection. Operates as a virtual COM port |
Ethernet (LAN) | Standard 10/100 base-T hardware connection. ICMP and TCP/IP Protocol for connection to Local Area Network or direct connection to a single PC |
GPIB (only with Option G) | The interface conforms with IEEE-488.1 and IEEE-488.2 |
LXI Compliance | LAN interface is compliant with LXI-C. (LXI is the abbreviation for LAN eXtensions for Instrumentation) |
Digital Programming Performance | |
Voltage Setting Resolution | 10mV |
Voltage Setting Accuracy | 0.05% +50mV |
Current Setting Resolution | 0.1mA (0.01mA on 75mA range) |
Current Setting Accuracy | ± (0.3% +0.1mA); ± (0.3% +0.01mA) on 75mA range |
Programming Speed | |
Command Delay | Typically <80ms (this must be added to any of the figures below) |
Voltage Up Time | Typically <45ms* to 1% |
Voltage Down Time | Typically <20ms* to 1% (full load); typically <150ms* to 1% (no load).. |